V1060:日本日立,雙蹤雙掃100M   銷售價格:800元  銷售熱線:010-67131181 /64021565/ 13601086585/ 歡迎有意者來電垂詢! 


V1060示波器 ? 的詳細介紹

X軸自動設業,CRT數字顯示,Y軸靈敏度:2mv/div-5v/div.X軸掃速:50ns-0.5s/div,有X10擴展,示波管為15cm方屏內刻度。外形尺寸:280X130X350mm,重約:5.5kg ,220v/35w

The Hitachi portable read-out oscilloscope provides measure-ment reliability and ease of operation by employing amicroprocessor.
The major features are:
(1) Measurement information display
The measurement information including the sweep speed, thedelay time, the IJNCAL display, and the voltage sensitivity ofthe vertical axis (V-1065 ,V-1065C and V-665 only) is alpha-numerically displayed on the CRT.
Since troublesome setting operation procedures are eliminat-ed, an operator can concentrate on the displayed data formeasurement.
(2) Measurement value display (V-1065, V-1065C, V-665 only)
The distance between the two cursors displayed can read out the following;
△V : Voltage between the reference cursor and the deltacursor
△T : Time between the reference cursor and the deltacursor
1/△T: Reciprocal of the time
The displayed data eliminates troublesome and time consumingcalculation procedures. Moreover, miscalculation of thescales is completely avoided.
(3) Automatic time base range setting
At a press of the AUTO button, an optimum time base range is automatically set.(A signal period from 1.6 to 4 cyclesapprox. is displayed.)The time base range is automatically changed to a correspond-ing change in input signal period.

(4) Trigger lock
Since complicated pulse train waveforms are hard to trigger on, performing the trigger lock function enables sweep independent triggering. The "sweep time plus holdoff time"is fixed and a stable trigger is obtained at any time range.

(5) Bandwidth
V-1065,V-1065C,: DC-100 MHz from 5 V/div to S mv/div
V-1060  DC-20 MHz at 2 mV/div
V-665, V-660 : DC-60 MHz from 5 V/div to S mv/div
(6) High sensitivity
DC-10 MHz at 2 mv/divHigh sensitivity of 2 mV/div is provided.

(7) Internal graticule
Internal graticule lines eliminate parallax-viewing errorbetween the trace and the graticule lines.
(8) Delayed sweep
With delayed sweep, a portion of the signal can be magnifined for more accurate measurement and the windows of delayed sweep(B) is indicated on the main sweep(A) with cursors.
(9)Auto trigger level
Auto measuring of trigger level is employed, so that trigger level range is mathed to the trigger signal for maximum trigger sensitivity and stability.
(10)TV triggering
Exclusive TV sync separator circuit technology provides stable TV signal measurements on fields, frames and lines.

Specifications
(A) Frequency counter
1.1 Specifications
Measuring signal : Measure the signal selected as atrigger signal by the TRIG SOURCEOR X switch.
Measuring range : 20 Hz to 100 MHz
Accuracy of reference signal : 100 ppM (15 to 35°C)

Measuring range Display format Resolution Accuracy
20Hz≤f<100Hz 99.99Hz 0.01Hz Ref.signal ±1 LSD
100Hz≤f<1kHz 999.9Hz 1.0Hz Max
1kHz≤f<10kHz 9.999Hz 0.002Hz Max
10kHz≤f<100kHz 99.99Hz 0.04Hz Max
100kHz≤f<1MHz 999.9Hz 0.1Hz
1MHz≤f<10MHz 9.999Hz 0.002Hz Max
10MHz≤f<100MHz 99.99Hz 0.01Hz
100MHz≤f (999.9Hz) Not specified

1 .2 Frequency measurement
  When MEASURE is selected by the SELECTOR switch, the AV cursor is displayed on CRT. Then press the SELECTOR switchthree times to display FREQ at the upper part of CRT.
  The frequency counter is interlocked with the triggering andthe measurement starts from the triggering point.
  Consequently, any special frequency having complex waveformscan be measured by changing the trigger level.
  For example, when a signal as illustrated below is applied,the frequency of the entire waveforms can be measured whentriggered at point A. When triggered at point B, thefrequency of a pulse with large amplitude can be measured.

Note : When not triggered, the measurement can not be per-formed, since the counter is interlocked with thetriggering. In the non-triggering mode or in case ofthe input signal less than 20 Hz, the message"FREQ..." is displayed.

(B) Memory backup
The panel setting data is retained for move than 48 hoursafter turning off power. It is not needed to perform thesame panel settings again even on the day after tomorrow.

(C) Other specifications
Same with those of the V-1065.

V-1065/V-1065C V-1060 V-665 V-660
●CRT
Graticule 6-inch screen with internal graticule 0%, 10%, 90% and 100% markers
8x10 DIV(1DIV=1cm)
Phosphor P31
Accelerating potential 17kV approx. 12kV approx.
External intensity modulation Coupling: DC coupling
Voltage: 5V or more
Maximum input voltage: 30V (DC+AC peak)
            or 30Vp-p AC at 1kHz or less
Bandwidth: DC to 5MHz
●Vertical deflection system
Sensitivity 2mV/DIV to 5V/DIV ±3%(switchable in 11 steps) Continuously variable
Bandwidth DC to 100MHz -3dB
2mV/DIV: DC to 20MHz -3dB
AC low pass: 10Hz
60MHz
10MHz
Rise time 3.5 ns apporx.
2mV/DIV: 17.5 ns approx.
5.9ns approx.
Delay time Leading edge can be monitoed 35ns approx.
Maximum input voltage 400V(DC+AC peak)at 1 kHz or less
Input coupling AC, DC, GND
Input impedance 1M ohms ±1.5%, 23pF ±3pF
Display modes CH1, CH2, DUAL, CHOP(250kHz approx.),ADD(DIFF mode can be established when the CH2 is in the INVERT mode.)
Bandwidth limiting function 20MHz 10MHz
Polarity selection ±(CH2 only)
Common-mode rejection ratio 20dB minimum at 20MHz
X-Y operation X-axis, Y-axis selectable
Sensitivity X axis:
CH1, CH2 2mV to 5V/DIV ±3%
EXT      0.1V/DIV ±5%
EXT÷10   1V/DIV ±5%
Y axis: 2mV to 5V/DIV ±5%
Phase error 3°or less from DC to 50 kHz
X bandwidth DC to 500 kHz(-3dB)
●Horizontal deflection system
Trigger mode Trigger,auto trigger
Sweep mode Main sweep, continous delay sweep, alternate sweep, signal sweep
Trigger source CH1, CH2, EXT(AC,DC,DC÷10),LINE
TV trigger Excusive sync separtor circuit provided Sync polarity:-
Trigger Sensitivity
NORM mode
Frequence DC to 20MHz 20MHz to 100MHz
INT 0.35 DIV 1.5 DIV
EXT 50 mV 150 mV
DC to 10MHz
10MHz-60MHz
AUTO mode
Frequence 30 to 100Hz 100Hz to 20MHz 20 to 100MHz
INT 1.5 DIV 1 DIV 1.5 DIV
EXT 150 mV 100 mV 150 mV
30 to 100Hz
100Hz-10MHz
10-60MHz
TV mode Sync signal
Internal sync: 1div or more
External sync: 200mVp-p or more
Trigger level Variable range
AUTO: Automatically corresponds to the trigger signal
NORM: INT: ±4 DIV or more
      EXT: ±0.4 V or more
      EXT÷10:±4V or more
External input impedance 1 M ohms ±5%, 25pF + 6pF
External input voltage 400V(DC+AC peak) at 1kHz
Sweep time
A(main) sweep 50 ns/DIV to 0.5 s/DIV   Continuously variable (UNCAL)
B(delay) sweep 50 ns/DIV to 50 ms/DIV
Accuracy
10 to 35℃ 0 to 50℃
X1 ±3% ±4%
MAG X10 ±4% ±6%
Sweep magnification X10
Maximum sweep rate 5 ns/DIV
Hold off time Variable
Delay time 1μs to 5s
Delay jitter 1/20,000 or less
Alternate separation Variable
Trigger lock function Provided
Auto range function Provided
Singnal sweep function Provided
●Readout funtion
Pannel setting display Vertical axis: V/DIV(CH1,CH2),UNCAL,probe conversion
Sweep speed: S/DIV,UNCAL,MAG(converted value)
Other: Delay time
Not provided Same as V-1065 Not provided
●Cursor readout function Voltage difference △V: △-REF
Time difference △T: △-REF
Frequency 1/△T: △-REF
No cursor Same as V-1065 No cursor
●External output
TRIGGER SIGNAL OUT Output voltage: 25mV/DIV approx.
(Full scale on the CRT)
50-ohm termination
Frequency response: DC to 10MHz
Output impedance: 50 ohms approx.
Not provided Same as V-1065 Not provided
●Calibrator
Waveform 1kHz ±20%, square wave
Voltage 0.5V ±1%
●Power supply 90V to 250V AC, 48 to 440Hz, 40W approx.
●Other
Dimensions 275(W) x 130(H) x 360(D) mm
Weigth 6kg
EMI VDE0871, Category B